弹簧管式精密压力表及真空表检定规程(国外英文资料)
弹簧管式精密压力表及真空表检定规程(国外英文资料)
The information that is worth having
It comes from the general learning accumulation summary
There must be a problem
Please be critical!
Spring tube type precision pressure gauge and vacuum gauge calibration procedure
JJG 49-87 instead of JJG 49-72
Regulation of the Precise Bourdon Pressure Gauge and Vacuum Gauge
This verification procedure was approved by the national bureau of measurement on April 11, 1987
It has been in effect since October 1, 1987
Unit: Shanghai standard measurement bureau
Drafting unit: the metrological management bureau of Shanghai standard metrology authority
The technical provisions of this discipline shall be interpreted by the drafting unit
The main drafters of this discipline:
Xujiarong (bureau of metrological management)
Spring tube type precision pressure gauge and vacuum gauge calibration procedure
This procedure applies to new manufacturing, use and repair of measurement limit of 0.1 ~ 250 mpa spring tube type precision pressure gauge and vacuum gauge verification (hereinafter referred to as the precision table)
Technical requirement
1 precision table Pointers (including replacement Pointers) should be within 0.5 to 1.5 mm range of surface. Its blade is perpendicular to the surface of the dial
The pointer indicates that the end should be reached in the shortest dividing line 1/4 ~ 3/4
The end width of the pointer blade should not be greater than the dividing line width
Precision watch dial and surface glass should not interfere with reading defects and damage
The zero position of the dial is not allowed to have a stop pin
The dial width should not be greater than 0.2 mm
The precision of 3000 points should be evenly spaced
The full length of the outer arc of the graduated ruler should meet the provisions of table 1
Table 1
The movement of the precise table pointer should be stable throughout the range
No jumping or jamming
The sealing device of the precision watch
In case of no damage to seal
Cant touch the i
您可能关注的文档
- 工业设计流程(国外英文资料).doc
- 工作中遇到的问题及小常识span(国外英文资料).doc
- 工作岗位评估的方法(国外英文资料).doc
- 工作原理和分类(国外英文资料).doc
- 工业润滑油使用中常见问题(国外英文资料).doc
- 工作总结2003年工作总结(电力2)(国外英文资料).doc
- 工作总结供电所2004年终工作总结(国外英文资料).doc
- 工作总结和工作设想(国外英文资料).doc
- 嵌入式(国外英文资料).doc
- 工作总结某县地方税务局2002年工作总结(国外英文资料).doc
- 《GB/T 32151.42-2024温室气体排放核算与报告要求 第42部分:铜冶炼企业》.pdf
- GB/T 32151.42-2024温室气体排放核算与报告要求 第42部分:铜冶炼企业.pdf
- GB/T 38048.6-2024表面清洁器具 第6部分:家用和类似用途湿式硬地面清洁器具 性能测试方法.pdf
- 中国国家标准 GB/T 38048.6-2024表面清洁器具 第6部分:家用和类似用途湿式硬地面清洁器具 性能测试方法.pdf
- 《GB/T 38048.6-2024表面清洁器具 第6部分:家用和类似用途湿式硬地面清洁器具 性能测试方法》.pdf
- 《GB/T 18238.2-2024网络安全技术 杂凑函数 第2部分:采用分组密码的杂凑函数》.pdf
- GB/T 18238.2-2024网络安全技术 杂凑函数 第2部分:采用分组密码的杂凑函数.pdf
- 《GB/T 17215.686-2024电测量数据交换 DLMS/COSEM组件 第86部分:社区网络高速PLCISO/IEC 12139-1配置》.pdf
- GB/T 13542.4-2024电气绝缘用薄膜 第4部分:聚酯薄膜.pdf
- 《GB/T 13542.4-2024电气绝缘用薄膜 第4部分:聚酯薄膜》.pdf
文档评论(0)